iChem VELOCITY Urine Chemistry Analyzer
Recall
- Recall Number
- Z-0843-2020
- Event Number
- 84268
- Firm
- Beckman Coulter Inc.
- FEI Number
- 2050012
- Product Code
- KQO
- Status
- Terminated
- Root Cause
- Error in labeling
- Initiated
- November 7, 2019
- Terminated
- September 11, 2023
- Address
- 250 S Kraemer Blvd, Brea, CA, 92821-6232
Description
iChem VELOCITY Urine Chemistry Analyzer
Beckman Coulter has become aware of an issue affecting Specific Gravity chemistry settings for the iChemVELOCITY Urine Chemistry System that prevents the trigger of a chemistry system flag S3 and may result in erroneous Specific Gravity (SG) results.
The firm, Beckman Coulter, sent an "URGENT MEDICAL DEVICE RECALL" letters on 11/07/2019. The letter explained the problem and how to recognize if the S3 flag has been disabled in the instrument. The letter further instructs the following: - Review results to determine if samples report an SG result of >1.060 or an SG result of S3 - If the SG result is >1.060 or S3, test samples for SG using an alternate method* determined by your laboratory - If persistent SG results are >1.060 or persistent S3 results are seen, contact your local Beckman Coulter Representative If you have any questions regarding this notice, please contact Beckman Coulter Customer Support Center: Via our website: http://www.beckmancoulter.com Support > Request Instrument Support. By phone: call 800-526-7694 in the United States and Canada. Outside the United States and Canada, contact your local Beckman Coulter Representative.
Worldwide Distribution: US (nationwide) including Puerto Rico and countries of: Algeria, Argentina, Australia, Austria, Bahrain, Bangladesh, Belarus, Belgium, Bolivia, Bosnia and Herzegovina, Brazil, Brunei, Bulgaria, Canada, Chile, China, Colombia, Croatia, Cyprus, Czech Republic, Ecuador, France, Germany, Ghana, Greece, Guatemala, Hong Kong, Hungary, India, Indonesia, Israel, Italy, Japan, Jordan, Kazakhstan, Kuwait, Lebanon, Libya, Macau, Malaysia, Maldives, Mexico, Moldova, Montenegro, Morocco, Netherlands, Netherlands Antillean, New Zealand, Nigeria, Oman, Pakistan, Panama, Peru, Philippines, Poland, Portugal, Qatar, Romania, Russia, Saudi Arabia, Serbia, Singapore, Slovakia, South Africa, South Korea, Spain, Switzerland, Taiwan, Thailand, Turkey, Uganda, Ukraine, United Arab Emirates, United Kingdom, Uzbekistan, Venezuela, Vietnam.
4314 units