FDA Recall Terminated

Simplastin HTF Reagents, 20ml HO, ( and 6 ml HO) bioMerieux, Inc., Durham, NC 27704

Recall: Z-0843-05 · Initiated March 7, 2005

Recall

Recall Number
Z-0843-05
Event Number
31482
Firm
BIOMERIEUX, INC.
FEI Number
3002769706
Product Code
GJS
Status
Terminated
Root Cause
Other
Initiated
March 7, 2005
Posted
May 18, 2005
Terminated
October 16, 2006
Address
100 Rodolphe St, Durham, NC, 27712-9402

Description

Simplastin HTF Reagents, 20ml HO, ( and 6 ml HO) bioMerieux, Inc., Durham, NC 27704

Reason

The ISI labeling for Simplastin HTF on the Coag-a-Mate MAX and Thrombolyzer instrument platforms is not correct. The Simplastin HTF ISI assignment was mis-assigned and the published values are higher than expected for the listed instrument platforms.

Action

Consignees were notified by letter on/about March 7, 2005.

Distribution

Nationwide, Canada, Colombia, India, Philippines, Spain, Thailand

Quantity

19748 kits