FDA Recall
Completed
Augustine Surgical Inc. HOT DOG PATIENT WARMING, Temperature Management Controller, Single Port, Models: 1) WC71V; 2) WC71V-DEMO; 3) WC71V-R; used in conjunction with Warming Blankets and Mattresses, veterinary use.
Recall: Z-0842-2025
·
Initiated March 14, 2022
Recall
- Recall Number
- Z-0842-2025
- Event Number
- 95913
- Firm
- Augustine Temperature Management, LLC
- FEI Number
- 3005857264
- Product Code
- DWJ
- Status
- Completed
- Root Cause
- Device Design
- Initiated
- March 14, 2022
- Posted
- January 10, 2025
- Address
- 15305 Minnetonka Blvd, Minnetonka, MN, 55345-1512
Description
Augustine Surgical Inc. HOT DOG PATIENT WARMING, Temperature Management Controller, Single Port, Models: 1) WC71V; 2) WC71V-DEMO; 3) WC71V-R; used in conjunction with Warming Blankets and Mattresses, veterinary use.
Reason
HotDog Temperature Management Controllers Models WC71, WC77, and Temperature Management Controller MP Model 2083516 may experience a Power-on self-test error code "EA POST".
Action
Augustine Temperature Management, LLC began correcting (installing ferrites and software updates) the affected devices February, 2022 and completed the actions in August 2023. The firm did not perform any notification to its consignees that corrections would be implemented.
Distribution
Worldwide
Quantity
2271 units