FDA Recall Terminated

GE OEC Brivo 865 Plus Mobile C-Arm X-Ray Product Used for general surgical applications and musculoskeletal procedures to visualize.

Recall: Z-0841-2015 · Initiated September 15, 2014

Recall

Recall Number
Z-0841-2015
Event Number
69660
Firm
GE OEC Medical Systems, Inc
FEI Number
1720753
Product Code
OXO
Status
Terminated
Root Cause
Radiation Control for Health and Safety Act
Initiated
September 15, 2014
Posted
January 13, 2015
Terminated
May 20, 2015
Address
384 N Wright Brothers Dr, Salt Lake City, UT, 84116-2862

Description

GE OEC Brivo 865 Plus Mobile C-Arm X-Ray Product Used for general surgical applications and musculoskeletal procedures to visualize.

Reason

mA Accuracy may exceed the design and labeling specification of+ 10%. Per 1020.32(f): Deviation of x-ray tube potential and current from the indicated values shall not exceed the maximum deviation as stated by the manufacturer in accordance with 1020.30(h)(3).

Action

Planned action(s) to repair defect or to bring product into compliance: GE Healthcare will, without charge, perform a mA calibration on all affected systems in the USA to bring the product back into compliance. The details of which will be included in a subsequent communication to customers or through a GE Healthcare field engineer site visit. Should customers have any questions or concerns regarding these corrections, please do not hesitate to contact the service team for further information at 800-874-7378 option 8.

Distribution

Worldwide Distribution

Quantity

61 units installed in US