FDA Recall Terminated

Elasto-Gel Wound Dressings, 3 product codes: DR8050, DR 8000, and DR 8000LV (for foreign distribution only), 4'' x 4'', Sterile, each package contains 1 wound dressing, 5 packages per intermediate box, 20 intermediate boxes (100 pkgs) per case. The label shows the product is Manufactured by Southwest Technologies, Inc., North Kansas City, MO.

Recall: Z-0841-06 · Initiated March 27, 2006

Recall

Recall Number
Z-0841-06
Event Number
35051
Firm
Southwest Technologies Inc
FEI Number
1929833
Product Code
FRO
Status
Terminated
Root Cause
Other
Initiated
March 27, 2006
Posted
May 4, 2006
Terminated
December 12, 2007
Address
1746 E Levee St N, Kansas City, MO, 64116-4404

Description

Elasto-Gel Wound Dressings, 3 product codes: DR8050, DR 8000, and DR 8000LV (for foreign distribution only), 4'' x 4'', Sterile, each package contains 1 wound dressing, 5 packages per intermediate box, 20 intermediate boxes (100 pkgs) per case. The label shows the product is Manufactured by Southwest Technologies, Inc., North Kansas City, MO.

Reason

The product is labeled as sterile and one of the lots failed sterility during the quarterly radiation dose audit.

Action

Recall letters were sent to all direct accounts via certified mail on 3/27/2006. The consignees were requested to return any affected product on hand and if the product was further distributed, notify their customers.

Distribution

Product was distributed to direct accounts in the United States and Belgium.

Quantity

90 cases of Product DR8000; 56 cases or Product DR8000LV; and 28 cases of Product DR8050. Each case contained 100 units.