FDA Recall Terminated

Randox CALCIUM (Ca) COLORIMETRIC METHOD RX DAYTONA PLUS Cat. No. CA 8309 R1, Arsenazo Reagent 4 x 20 ml GTIN: 05055273208368

Recall: Z-0840-2019 · Initiated December 4, 2018

Recall

Recall Number
Z-0840-2019
Event Number
81746
Firm
Randox Laboratories Ltd. 55 Diamond Road Crumlin (North) Ireland
FEI Number
1000361607
Product Code
CJY
Status
Terminated
Root Cause
Device Design
Initiated
December 4, 2018
Terminated
June 18, 2020

Description

Randox CALCIUM (Ca) COLORIMETRIC METHOD RX DAYTONA PLUS Cat. No. CA 8309 R1, Arsenazo Reagent 4 x 20 ml GTIN: 05055273208368

Reason

Update to the carryover avoidance technical bulletin to introduce additional steps for reagent carryover avoidance with the Calcium assay on RX instruments. Carryover to the Calcium reagent results may result in falsely elevated or decreased results to both patient and QC samples.

Action

Randox the US distributor contacted customers by email on 12/4/18 to advise of problem and provided an update to the carryover avoidance technical bulletin to introduce additional steps for reagent carryover avoidance with the Calcium assay on RX instruments.The instrument testing order should be reviewed in line with the updated technical bulletin. A revised letter issued 12 Feb 2019 to advise "reagent would be observed as inconsistencies in Quality control recovery which may lead to a delay in running patient samples or erroneous elevated / depressed test results".

Distribution

WV, PR

Quantity

1 kit