FDA Recall
Terminated
Cook Urological Endoureterotomy Stent Set, 10Fr/13 cm, 6.0Fr/20 cm, sterile, Cook Urological, Inc., Spencer, IN; Ref EUSCFS-060020-10-13; GPN REF G18117. Ureteral stent.
Recall: Z-0840-2009
·
Initiated July 31, 2008
Recall
- Recall Number
- Z-0840-2009
- Event Number
- 48760
- Firm
- Vance Products Incorporated
- FEI Number
- 1825146
- Product Code
- FAD
- Status
- Terminated
- Root Cause
- Process control
- Initiated
- July 31, 2008
- Posted
- January 23, 2009
- Terminated
- November 27, 2009
- Address
- 1100 W Morgan St, Spencer, IN, 47460-9426
Description
Cook Urological Endoureterotomy Stent Set, 10Fr/13 cm, 6.0Fr/20 cm, sterile, Cook Urological, Inc., Spencer, IN; Ref EUSCFS-060020-10-13; GPN REF G18117. Ureteral stent.
Reason
The bond may break where the two materials meet, resulting in separation of the stent.
Action
Consignees were notified via "Urgent: Recall Notice" letter dated 7/31/08. The letter described the failure, instructed that the products be isolated and returned to the firm. For additional information, contact Cook Urological at 1-812-339-2235, ext. 7241.
Distribution
Nationwide, Australia, Canada, Colombia, Hong Kong, India, Ireland, Israel, Japan, Malaysia, Mexico, South Korea, Taiwan, Venezuela and Viet Nam
Quantity
1739 of all products.