Smith & Nephew, 420XL Xenon Light Source, 180 Watt; for endoscopic use. Catalog/Part #72200510; manufactured for Smith & Nephew Endoscopy, Andover, MA 01810.
Recall
- Recall Number
- Z-0840-2007
- Event Number
- 37849
- Firm
- Smith & Nephew, Inc Endoscopy Division
- FEI Number
- 1643264
- Product Code
- FFS
- Status
- Terminated
- Root Cause
- Nonconforming Material/Component
- Initiated
- April 9, 2007
- Posted
- June 7, 2007
- Terminated
- September 2, 2008
- Address
- 76 S. Meridian Ave, Oklahoma City, OK, 73107-6512
Description
Smith & Nephew, 420XL Xenon Light Source, 180 Watt; for endoscopic use. Catalog/Part #72200510; manufactured for Smith & Nephew Endoscopy, Andover, MA 01810.
Several units were manufactured using a power supply that contains a 1 amp fuse where a 5 amp fuse is required. When the unit's power supply draws more that 1A, the fuse will fail causing the light to cease operating. The problem can occur anytime: from the initial turning on (powering up), to anytime during use.
Notification sent 04/09/07 via FedEx to distribution centers, sales reps and user facilities. Users asked to send light source to Smith & Nephew GMBH Service Center in Germany for repair (contact Ute Dieling at +49 (0) 7462-208-313).
Product sent to distributors in Argentina, Austria, Chile, Denmark, Dubai, France, Greece, Hong Kong, India, Italy, Japan, Korea, Malaysia, Norway, South Africa, Spain, Sweden and United Kingdom. No US distribution.
67 units.