FDA Recall Terminated

Dunlee DA 200 Ultra & MX 200 Ultra CT X-Ray Tube Housing Assembly (989600088701); replacement x-ray tube housing assemblies; Dunlee, 555 North Commerce Street, Aurora, Illinois 60504 U.S.A.; part number 875-01

Recall: Z-0840-05 · Initiated April 7, 2005

Recall

Recall Number
Z-0840-05
Event Number
31457
Firm
Dunlee
FEI Number
1483239
Product Code
ITY
Status
Terminated
Root Cause
Other
Initiated
April 7, 2005
Posted
May 13, 2005
Terminated
October 20, 2005
Address
555 Commerce St, Aurora, IL, 60504-8110

Description

Dunlee DA 200 Ultra & MX 200 Ultra CT X-Ray Tube Housing Assembly (989600088701); replacement x-ray tube housing assemblies; Dunlee, 555 North Commerce Street, Aurora, Illinois 60504 U.S.A.; part number 875-01

Reason

The x-ray tube housing assembly may have been mounted with the incorrect sized bolts.

Action

Dunlee sent recall letters dated 4/4/05 to end user locations, informing them that Dunlee and their service representatives would visit each location to replace the mounting bolts on all Dunlee DA-200 Performix installations because of problems experienced at certain GE CT installations. Any questions were directed to John M. Macko, Product Marketing Manager, at 630-585-2174.

Distribution

Nationwide and internationally to Taiwan, Germany and Mexico.

Quantity

37 units