Roller Pump, 4 Inch Diameter The Terumo Advanced Perfusion System 1 is indicated for use for up to 6 hours on the extracorporeal circulation of blood for arterial perfusion, regional perfusion, and cardiopulmonary bypass procedures, when used by a qualified medical professional who is experienced in the operation of this or similar equipment.

Recall: Z-0839-2012 · Initiated January 13, 2012

Recall

Recall Number
Z-0839-2012
Event Number
60463
Firm
Terumo Cardiovascular Systems Corporation
FEI Number
1828100
Product Code
DTQ
Status
Terminated
Root Cause
Nonconforming Material/Component
Initiated
January 13, 2012
Posted
January 24, 2012
Terminated
December 3, 2012
Address
6200 Jackson Road, Ann Arbor, MI, 48103-9586

Description

Roller Pump, 4 Inch Diameter The Terumo Advanced Perfusion System 1 is indicated for use for up to 6 hours on the extracorporeal circulation of blood for arterial perfusion, regional perfusion, and cardiopulmonary bypass procedures, when used by a qualified medical professional who is experienced in the operation of this or similar equipment.

Reason

Terumo Cardiovascular Systems (Terumo CVS) discovered and rejected a single digital encoder cable assembly during incoming inspection because it had a loose wire. The assembly is used in the 4" and the 6" roller pumps for the Terumo Advanced Perfusion System 1.

Action

TERUMO sent an Urgent Medical Device Correction letter dated January 13, 2012, to all affected consignees. The letter identified the product, the problem, and the action to be taken by the consignee. Consignees were instructed to review the Medical Device Correction notice. Assure that all users are aware of the notice. Confirm receipt of the communication by faxing the attached Customer Response Form to 1-900-292-6551. Roller pumps with affected cables will have the cables assembly replaced. For questions regarding this recall call 1-800-521-2818.

Distribution

Worldwide Distribution - USA including MD, WI, and IA and the countries of Canada, Mexico, and Vietnam

Quantity

44