Outlook 400ES Safety Infusion System, Model Number 621-400ES. For use with B. Braun Medical Inc. Horizon Pump IV Sets
Recall: Z-0837-2011
·
Initiated August 24, 2010
Recall
- Recall Number
- Z-0837-2011
- Event Number
- 57099
- Firm
- B Braun Medical, Inc
- FEI Number
- 1641965
- Product Code
- FRN
- Status
- Terminated
- Root Cause
- Component design/selection
- Initiated
- August 24, 2010
- Posted
- January 10, 2011
- Terminated
- May 29, 2014
- Address
- 1601 Wallace Dr, Ste 150, Carrollton, TX, 75006-6690
Description
Outlook 400ES Safety Infusion System, Model Number 621-400ES. For use with B. Braun Medical Inc. Horizon Pump IV Sets
Reason
Possible pump stop with audible alarm, but no visual error indication that the pump is not delivering the medication. Under certain conditions in a network environment, Management Processor may become non-responsive, ceasing normal operation.
Action
B. Braun Medical Inc. notified consignee by letter on August 24, 2010.
Distribution
Virginia and New York
Quantity
2,107 Pumps (582 Pumps to Winchester)