FDA Recall Terminated

Outlook 400ES Safety Infusion System, Model Number 621-400ES. For use with B. Braun Medical Inc. Horizon Pump IV Sets

Recall: Z-0837-2011 · Initiated August 24, 2010

Recall

Recall Number
Z-0837-2011
Event Number
57099
Firm
B Braun Medical, Inc
FEI Number
1641965
Product Code
FRN
Status
Terminated
Root Cause
Component design/selection
Initiated
August 24, 2010
Posted
January 10, 2011
Terminated
May 29, 2014
Address
1601 Wallace Dr, Ste 150, Carrollton, TX, 75006-6690

Description

Outlook 400ES Safety Infusion System, Model Number 621-400ES. For use with B. Braun Medical Inc. Horizon Pump IV Sets

Reason

Possible pump stop with audible alarm, but no visual error indication that the pump is not delivering the medication. Under certain conditions in a network environment, Management Processor may become non-responsive, ceasing normal operation.

Action

B. Braun Medical Inc. notified consignee by letter on August 24, 2010.

Distribution

Virginia and New York

Quantity

2,107 Pumps (582 Pumps to Winchester)