FDA Recall Terminated

LIFEPAK EXPRESS defibrillator

Recall: Z-0837-2007 · Initiated October 5, 2006

Recall

Recall Number
Z-0837-2007
Event Number
36484
Firm
Medtronic Emergency Response Systems, Inc.
FEI Number
3015876
Product Code
MKJ
Status
Terminated
Root Cause
Nonconforming Material/Component
Initiated
October 5, 2006
Posted
June 21, 2007
Terminated
August 8, 2013
Address
11811 Willows Rd Ne, Redmond, WA, 98052-2003

Description

LIFEPAK EXPRESS defibrillator

Reason

Premature battery depletion due to electrical component failure

Action

Per request from Japan, the firm initiated action on 10/5/06 replacing the devices in Japan with new devices. On 3/2/07 the firm issued a Safety Alert letter dated February 2007 to consignees (with the exception of Japan), advising that the defibrillators have an identified, lead-free component that is susceptible to failure within the first 250 days after manufacture. Failure of this component might cause a short that will drain the battery and eventually render the device inoperative. The letter further states that the firm strongly recommends that the consignee immediately and on a regular basis, inspect their defibrillator(s) per the LIFEPAK CR Plus/EXPRESS Operating Instructions - Section 5 Caring for the Defibrillator. The letter also provides a website to allow consignees to enter their device serial number(s) to determine if their LIFEPAK CR Plus/LIFEPAK EXPRESS are included in the action. This recall strategy is subject to modification pending the performance of devices remaining on the market.

Distribution

Distribution is worldwide. The device is designed for non-professional users and are primarily used in public access defibrillator environments such as airports, schools, and public libraries.

Quantity

Total of Express 1,006 units worldwide; amount does not include units distributed to Japan.