DeRoyal (R) Aquarius Hot/Cold Therapy Combo Unit, w/ Knee/Shoulder Blanket NS, REF T652NS, Rx only, Manufacturer DeRoyal Industries, Inc. 200 DeBusk Lane, Powell, TN 37849. Physical Therapy.
Recall
- Recall Number
- Z-0835-2014
- Event Number
- 66742
- Firm
- DeRoyal Industries Inc
- FEI Number
- 3003040625
- Product Code
- ILO
- Status
- Terminated
- Root Cause
- Device Design
- Initiated
- September 17, 2013
- Posted
- January 27, 2014
- Terminated
- June 5, 2017
- Address
- 1595 Highway 33 S, New Tazewell, TN, 37825-7105
Description
DeRoyal (R) Aquarius Hot/Cold Therapy Combo Unit, w/ Knee/Shoulder Blanket NS, REF T652NS, Rx only, Manufacturer DeRoyal Industries, Inc. 200 DeBusk Lane, Powell, TN 37849. Physical Therapy.
The unit displays an error code of hose kinking when no hose kink is present; the unit fails to monitor the temperature of the water in the blanket as intended and controlling it through the automatic temperature control. This failure could result in temperature injuries to the treatment site; and there have been two reports of electric shock.
DeRoyal sent a "URGENT VOLUNTARY RECALL DEROYAL HOT/COLD THERAPY UNIT" letter dated September 17, 2013. The letter identified the product, problem, and actions to be taken by the customers.
Nationwide Distribution.
121 units