FDA Recall Terminated

Fluorescence Imaging Procedure Kit, designed for use with Firefly Fluorescence Imaging for the da Vinci Si 3000 Surgical Systems. The Intuitive Surgical Endoscopic Instrument Control System (Intuitive Surgical da Vinci Surgical System Model IS3000) is intended to assist in the accurate control of Intuitive Surgical Endoscopic Instruments.

Recall: Z-0834-2015 · Initiated November 21, 2014

Recall

Recall Number
Z-0834-2015
Event Number
69845
Firm
Intuitive Surgical, Inc.
FEI Number
3001675293
Product Code
GCJ
Status
Terminated
Root Cause
Labeling design
Initiated
November 21, 2014
Posted
December 19, 2014
Terminated
February 20, 2015
Address
1266 Kifer Rd, Bldg 100, Sunnyvale, CA, 94086-5304

Description

Fluorescence Imaging Procedure Kit, designed for use with Firefly Fluorescence Imaging for the da Vinci Si 3000 Surgical Systems. The Intuitive Surgical Endoscopic Instrument Control System (Intuitive Surgical da Vinci Surgical System Model IS3000) is intended to assist in the accurate control of Intuitive Surgical Endoscopic Instruments.

Reason

The storage conditions for indocyanine green (ICG) are not on the individual Fluorescence Imaging Procedure Kit box, the six-kit box, and the shipping box. If heat were to damage the material, the result would be a loss of fluorescence.

Action

Urgent Medical Device Correction Letters were sent to customers on November 24, 2014.

Distribution

Nationwide Distribution.

Quantity

1532 total both devices