FDA Recall Terminated

HENRY SCHEIN(R) 100-8008 Explorer-Double End #5. Barcode *+H658100800808*. Made in China. For professional use only. Instruments must be cleaned and sterilized prior to use. Distributed by: HENRY SCHEIN, INC. 135 Duryea Road, Melville, NY 11747 USA. This product is a dental explorer made of stainless steel packaged individually within a sleeve and containing a cardboard insert (labeling). 510 (k) exempt, Class I device, Device Listing # B057701.

Recall: Z-0834-04 · Initiated March 29, 2004

Recall

Recall Number
Z-0834-04
Event Number
28672
Firm
Hu-Friedy Mfg Co, Inc.
FEI Number
1416605
Product Code
EKB
Status
Terminated
Root Cause
Other
Initiated
March 29, 2004
Posted
July 20, 2004
Terminated
August 2, 2012
Address
3232 N Rockwell St, Chicago, IL, 60618-5944

Description

HENRY SCHEIN(R) 100-8008 Explorer-Double End #5. Barcode *+H658100800808*. Made in China. For professional use only. Instruments must be cleaned and sterilized prior to use. Distributed by: HENRY SCHEIN, INC. 135 Duryea Road, Melville, NY 11747 USA. This product is a dental explorer made of stainless steel packaged individually within a sleeve and containing a cardboard insert (labeling). 510 (k) exempt, Class I device, Device Listing # B057701.

Reason

Reports of breakage on some dental explorers were received by the own label distributor and forwarded to Hu-Friedy Mfg. Co. (importer of record & Spec. Setter). Investigation conducted by Hu-Friedy Mfg. Co. revealed that the tips of the instruments may have been improperly heat treated and have the potential for unexpected tip fracture.

Action

On 3/29/04, Hu-Friedy Mfg. Co. notified Henry Schein, Inc. of the recall by telephone. This was followed up with a letter on 4/2/04. Henry Schein initiated a sub-recall to its customers by sending recall letters and response forms (dated 3/31/04) by First Class Mail on 4/2/04 & 4/5/04. The mailing will be followed up with fax, e-mail or phone calls.

Distribution

The products were sold to the sole direct account: the Own Label Distributor, Henry Schein, Inc. in Melville, NY. Henry Schein further distributed the explorers to end users nationwide and internationally. Henry Schein Divisions located in Canada, Spain, France, Germany, United Kingdom, Holland, and New Zealand who further distributed the explorers to their customers in Canada and abroad.