FDA Recall Terminated

Centricity Enterprise Archive is a software product for receiving, archiving and sending of medical data. DICOM devices (e..g. modalities, workstations) Communicate with the archive using the DICOM protocol (published by ACR-NEMA). The clinical use is limited to the interaction that other systems have with the Centricity Enterprise archive. The intended users for Centricity Enterprise Archive are service personnel and system administrators and non clinical users. As of V4.0 XDS enabled systems can communicate with the archive using the XDS and XDS-1 profiles (published by IHE).

Recall: Z-0833-2015 · Initiated April 19, 2013

Recall

Recall Number
Z-0833-2015
Event Number
64760
Firm
Ge Healthcare It
FEI Number
3004526608
Product Code
LMB
Status
Terminated
Root Cause
Software design
Initiated
April 19, 2013
Posted
December 19, 2014
Terminated
June 10, 2016
Address
540 W Northwest Hwy, Barrington, IL, 60010-3051

Description

Centricity Enterprise Archive is a software product for receiving, archiving and sending of medical data. DICOM devices (e..g. modalities, workstations) Communicate with the archive using the DICOM protocol (published by ACR-NEMA). The clinical use is limited to the interaction that other systems have with the Centricity Enterprise archive. The intended users for Centricity Enterprise Archive are service personnel and system administrators and non clinical users. As of V4.0 XDS enabled systems can communicate with the archive using the XDS and XDS-1 profiles (published by IHE).

Reason

GE Healthcare has become aware of a potential data loss issue associated with empty path names resulting from HL-7 updates when the cache disk is almost full.

Action

General Electric sent an Urgent Medical Device Correction letter dated April 19, 2013, to all affected customers. The letter instructs customers to ensure that Enterprise Archive partitions have enough free space available, in particular the cache disk should have at least 10% free space available. This could be facilitated by setting appropriate disk limits and watermarks on the individual libraries in Enterprise Archive. In case the disks are not being purged to the configured watermarks and the cache disk is almost full, customers are requested to contact their local GE service person. A permanent product correction will be provided to all customers under FMI85209R. Customers with questions were instructed to contact their local Sales/Service Representative. Also the Remote Online Center (ROC) Customer Care can be reached at 1-800-437-1171.

Distribution

Nationwide Distribution including AZ, OH, WA.

Quantity

115 units