FDA Recall Terminated

Urethral Warmer System Model number FPRCH3001 The Urethral Warming System is designed to circulate a warm solution through a warming catheter to maintain urethral tissue near body temperatures while the surrounding prostate tissue is being frozen. It is intended to warm urethral tissue when performing cryogenic destruction of prostatic tissue with a Galil Medical Cryoablation System.

Recall: Z-0832-2011 · Initiated November 11, 2010

Recall

Recall Number
Z-0832-2011
Event Number
57254
Firm
Galil Medical
Product Code
GEH
Status
Terminated
Root Cause
Other
Initiated
November 11, 2010
Posted
December 29, 2010
Terminated
December 10, 2011
Address
4364 Round Lake Road, Arden Hills, MN, 55112

Description

Urethral Warmer System Model number FPRCH3001 The Urethral Warming System is designed to circulate a warm solution through a warming catheter to maintain urethral tissue near body temperatures while the surrounding prostate tissue is being frozen. It is intended to warm urethral tissue when performing cryogenic destruction of prostatic tissue with a Galil Medical Cryoablation System.

Reason

This voluntary recall has been initiated because our supplier (Gaymar Industries, Inc.) notified us that these products contain power cords that may crack or fail at or inside the plug due to a prong design problem. Affected plugs can be identified by a round ground pin and black plastic bridges. Cords with plugs that are not affected have U-shaped ground pins and no plastic bridges. Although Gali

Action

Gaymar sent an URGENT: MEDICAL DEVICE RECALL letter dated September 30, 2010, to all consignees. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to: Customers were instructed to: " Inspect units in their possession to determine if they contain affected power cord plugs or if there were any signs of excessive wear or damage. " Immediately discontinue use and quarantine the affected product. " Contact Customer Service Department at 800-828-7341 to arrange for replacement. " Complete the attached form for each unit and follow the instructions given to return the form by fax or e-mail. " If customers further distributed the product to another person or entity, they were to identify their customers and notify them at once of this product recall by including a copy of the recall notification letter and the attached form. For questions regarding this recall call 800-828-7341 or 716-662-2551.

Distribution

Worldwide Distribution - USA, including CA, GA, TX and the country of Canada.

Quantity

4