Surgical Headlight labeled in part *Gulf Medical Fiberoptics*OLDSMAR, FLORIDA U.S.A.* Model Numbers: GMF-H001, GMF-H100, GMF-H101, GMF-H102, GMF-H103, GMF-H104, GMF-H105, and GMF-H106. The Surgical Headlight is a passive luminaire. It is illuminated by a source of light by way of the fiber optics and provides supplemental lighting for medical and surgical procedures.

Recall: Z-0831-2012 · Initiated May 13, 2011

Recall

Recall Number
Z-0831-2012
Event Number
58983
Firm
Gulf Medical Fiberoptics, Inc.
FEI Number
3003768921
Product Code
FST
Status
Terminated
Root Cause
Process control
Initiated
May 13, 2011
Posted
January 20, 2012
Terminated
August 28, 2012
Address
448 Commerce Blvd., Oldsmar, FL, 34677

Description

Surgical Headlight labeled in part *Gulf Medical Fiberoptics*OLDSMAR, FLORIDA U.S.A.* Model Numbers: GMF-H001, GMF-H100, GMF-H101, GMF-H102, GMF-H103, GMF-H104, GMF-H105, and GMF-H106. The Surgical Headlight is a passive luminaire. It is illuminated by a source of light by way of the fiber optics and provides supplemental lighting for medical and surgical procedures.

Reason

Gulf Medical Fiberoptics, Inc. recalled their Surgical Headlight because the cleaning instructions for the exterior surfaces of the headlight , headgear and fiber cable with 70% isopropyl alcohol had not been validated.

Action

On May 16, 2011, Gulf Medical Fiberoptics, Inc. issued to all customers an updated letter to their May 13, 2011 "Field Correction of Instructions For Use" informing them to discontinue the use of 70% isopropyl to clean the surfaces of the device. Enclosed with the letter was an updated Surgical Headlight User Guide with instructions to discard/shred/recycle the prior guide. The firm asked customers to return a form acknowledging the receipt and action from the letter. Customers with questions should contact Customer Service department manager Jacqueline Armand at 813-855-6618 ext. 427.

Distribution

Worldwide Distribution - USA, including the states of FL, WI, WA, AL, CA, MO, MI, OH, VA, CO, TX, TN, NC and the countries of Turkey, Brazil, Belgium, South Korea, Panama, Mexico, Ireland, Taiwan (ROC), Italy, Argentina, India, Spain, Germany, and Venezuela.

Quantity

381 units