FDA Recall Open, Classified

Siemens Dimension Vista Magnesium Flex reagent cartridge (MG) Material # 10445158, Catalog # K3057 used in the Dimension Vista System. The Dimension Vista Magnesium Flex Reagent (MG) method is an In vitro diagnostic test for the quantitative measurement of magnesium in human serum, plasma, and urine on the Dimension Vista System. Magnesium measurements are used in the diagnosis and treatment of hypomagnesemia (abnormally low plasma levels of magnesium) and hypomagnesemia (abnormally high plasma levels of magnesium).

Recall: Z-0830-2022 · Initiated January 28, 2022

Recall

Recall Number
Z-0830-2022
Event Number
89697
Firm
Siemens Healthcare Diagnostics, Inc.
FEI Number
2517506
Product Code
JGJ
Status
Open, Classified
Root Cause
Under Investigation by firm
Initiated
January 28, 2022
Address
500 Gbc Dr, Newark, DE, 19702-2466

Description

Siemens Dimension Vista Magnesium Flex reagent cartridge (MG) Material # 10445158, Catalog # K3057 used in the Dimension Vista System. The Dimension Vista Magnesium Flex Reagent (MG) method is an In vitro diagnostic test for the quantitative measurement of magnesium in human serum, plasma, and urine on the Dimension Vista System. Magnesium measurements are used in the diagnosis and treatment of hypomagnesemia (abnormally low plasma levels of magnesium) and hypomagnesemia (abnormally high plasma levels of magnesium).

Reason

Dimension Vista Magnesium Flex reagent cartridge lots exhibiting imprecision and intermittent Abnormal Assay flags. When this issue occurs, there is a potential for QC failures or erroneous patient results which may lead to repeat and follow-up testing.

Action

The firm, Siemens Healthineers, sent "URGENT MEDICAL DEVICE CORRECTION" (VC-22-01.A.US) on 01/28/2022 via Fed Ex to all customers to inform them of the issue and instruct them to discontinue use and discard the affected lots. Complete and return the Field Correction Effectiveness Check Form via email at [email protected] or fax to: (312) 275-7795. At the present time, Siemens has no replacement lots to offer customers. Siemens suggests using an alternate methodology for magnesium until a replacement lot is available. If you received any complaints of illness or adverse events associated with the products listed in Table 1, immediately contact your local Siemens Remote Services Center or your local Siemens Technical Support Representative. Replacement kits will be provided when a new lot is available. If you have any questions, please contact your Siemens Healthineers Remote Services Center or your local Siemens Healthineers Technical Support Representative.

Distribution

Worldwide distribution: US (nationwide) 47 States including: AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, TN, TX, UT, VA, WA, WI, WV, WY) and OUS countries of: Australia, Canada and Netherlands.

Quantity

14220 (US 13305, OUS 915)