FDA Recall
Terminated
ABG Replacement Pump Assembly Catalog Number: 37092
Recall: Z-0830-04
·
Initiated March 30, 2004
Recall
- Recall Number
- Z-0830-04
- Event Number
- 28725
- Firm
- Nova Biomedical Corporation
- FEI Number
- 1219029
- Product Code
- CHL
- Status
- Terminated
- Root Cause
- Other
- Initiated
- March 30, 2004
- Posted
- July 20, 2004
- Terminated
- June 17, 2004
- Address
- 200 Prospect St, Waltham, MA, 02453-3457
Description
ABG Replacement Pump Assembly Catalog Number: 37092
Reason
Incorrectly assembled pumps may result in inconsistent compression of the tubing causing erratic results or calibration problems
Action
Nova Biomedical notified users by telephone on 3/31/04 and conducted follow-up by letter either by email or fax. On-site visit will be scheduled to replace the pump assembly.
Distribution
CA, IL, LA, MA ,NJ, NY, OH, PA, TX International: Italy ,England, Japan,Turkey., U.A.E., Greece, Thailand, Switzerland, Finland, Saudi Arabia, Italy, Mexico, Trinidad, Ireland, France, Cyprus, Czech Republic, ,Germany
Quantity
5 units