Advisor Vital Signs Monitor (model 9200) with Capnography options installed (model 9212S or model 9212A), catalog no. 925574325.
Recall
- Recall Number
- Z-0828-05
- Event Number
- 31505
- Firm
- Smiths Medical PM, Inc.
- FEI Number
- 2183502
- Product Code
- DPS
- Status
- Terminated
- Root Cause
- Other
- Initiated
- March 16, 2005
- Posted
- May 12, 2005
- Terminated
- December 31, 2006
- Address
- N7w22025 Johnson Dr, Waukesha, WI, 53186-1856
Description
Advisor Vital Signs Monitor (model 9200) with Capnography options installed (model 9212S or model 9212A), catalog no. 925574325.
Altitude testing of the Advisor with Capnography module found the CO2 (End Tidal and Inspired) values are incorrect when the Advisor is configured to display in the relative units of percent (%) at elevations above 4,100 feet (at standard temperature pressure). The % CO values displayed will be lower than the actual CO2 value.
On March 16, 2005, customers were emailed Safety Action Bulletin 05-003 informing them of the low CO2 (End Tidal and Inspired) measurements in percent (%) and urging them to use mmHg or kPa as the unit of measurement . In addition, a Technical Worksheet will be distributed to customers who request a software upgrade. The Technical Worksheet will instruct them how to perform the software upgrade which corrects the CO2 measurements (End Tidal and Inspired) in percent (%).
CA and NY within US. Mexico, Argentina, Columbia, Chile, Brazil, Ecuador, Guatemala and Pakistan.
30