Evita Infinity V500 Ventilator; Babylog Infinity VN500 Ventilator. Ventilation.

Recall: Z-0827-2016 · Initiated February 4, 2016

Recall

Recall Number
Z-0827-2016
Event Number
73241
Firm
Draeger Medical, Inc.
FEI Number
2517967
Product Code
CBK
Status
Terminated
Root Cause
Software design
Initiated
February 4, 2016
Posted
February 25, 2016
Terminated
April 1, 2020
Address
3135 Quarry Rd, Telford, PA, 18969-1042

Description

Evita Infinity V500 Ventilator; Babylog Infinity VN500 Ventilator. Ventilation.

Reason

The battery capacity of the optional PS500 power supply unit for the Infinity (ACS) Workstation Critical and Neonatal Care was rapidly and unexpectedly reduced, despite using the most recent power supply firmware version 1.50 released in November 2015 (recall Z-0436-2016). The batteries were prematurely discharged, even though an adequate battery charge status was displayed on the device.

Action

Draeger Medical, Inc. will send an Urgent Medical Device Recall Letter to consignees in a method that confirms delivery. For consignees that have received new devices with firmware version 1.50 (127 devices) or through device updates (644 devices; reference Recall Z-0436-2016), those devices will be downgraded to firmware version 1.49 and the batteries replaced. Batteries will continue to be replaced for all consignees every six months until a final solution is available.

Distribution

US Nationwide Distribution including Puerto Rico.

Quantity

2553