Siemens syngo.plaza VA20A Server Farm setup integrated to a RIS, model number 10592457.

Recall: Z-0827-2012 · Initiated May 1, 2011

Recall

Recall Number
Z-0827-2012
Event Number
60761
Firm
Siemens Medical Solutions USA, Inc
FEI Number
2240869
Product Code
LLZ
Status
Terminated
Root Cause
Software design
Initiated
May 1, 2011
Posted
January 20, 2012
Terminated
January 24, 2014
Address
51 Valley Stream Pkwy, Malvern, PA, 19355

Description

Siemens syngo.plaza VA20A Server Farm setup integrated to a RIS, model number 10592457.

Reason

The firm has become aware of the potential for an unintended merge of multiple patients when using syngo.plaza version VA20 with Server Farm setup integrated to a RIS. Sporadically it can happen that certain Patient IDs in the database are deleted and sent to NULL. This can cause incorrect merging with patients that already had a NULL Patient ID.

Action

The firm initiated their product recall in May 2011 by issuing a "Customer Safety Advisory Notice" informing all customers of the malfunction and providing instructions on how to avoid the error. The issue will be solved with the next version release, planned for June 2011.

Distribution

Nationwide Distribution - Including the states of Florida, Massachusetts and Wisconsin.

Quantity

3