FDA Recall Terminated

4Z1c matrix array transducer which is used with ACUSON SC2000 ultrasound system. Siemens Medical Solutions USA, Inc. Business Unit Ultrasound 1230 Shorebird Way Mountain View, CA 94043 Intended use: 4Z1 transducer is a special purpose ultrasound transducer designed to help clinicians perform adult echocardiography and pre-operative and intra-operative surgical assessments. Secondary applications include: pediatric, echocardiography, primarily benefiting older and larger pediatric populations, general abdominal and Pinta-operative pediatric abdominal imaging and peripheral vessel imaging.

Recall: Z-0826-2011 · Initiated September 29, 2010

Recall

Recall Number
Z-0826-2011
Event Number
57289
Firm
Siemens Medical Solutions USA, Inc.
FEI Number
2936884
Product Code
IYN
Status
Terminated
Root Cause
Other
Initiated
September 29, 2010
Posted
December 28, 2010
Terminated
October 4, 2011
Address
1230 Shorebird Way P.O. Box 7393, Mountain View, CA, 94043

Description

4Z1c matrix array transducer which is used with ACUSON SC2000 ultrasound system. Siemens Medical Solutions USA, Inc. Business Unit Ultrasound 1230 Shorebird Way Mountain View, CA 94043 Intended use: 4Z1 transducer is a special purpose ultrasound transducer designed to help clinicians perform adult echocardiography and pre-operative and intra-operative surgical assessments. Secondary applications include: pediatric, echocardiography, primarily benefiting older and larger pediatric populations, general abdominal and Pinta-operative pediatric abdominal imaging and peripheral vessel imaging.

Reason

There is the possibility of a short occurring in the transducer which could cause focal heating on the probe face.

Action

SIEMENS issued a filed correction in the form of a "Customer Safety Advisory Notification" initiated on September 29, 2010 and mailed to customers by Federal Express on the same day. The notification described the product, problem and actions to be taken by the customers. The customers were instructed to examine the 4Z1c transducer prior to use (Note: if transducer does not show a blister on its front face or does not feel unusually when it is safe for use); to immediately discontinue use and contact their local Siemens representative to get issue resolved, if the transducers developed a hot spot that exhibits a visible blister on the patient contact surface and feels hot to the touch during operation, and to pass this notice on to all those within their organization who need to be aware. If you have any questions, please contact (650) 969-9112 or (800) 422-8766.

Distribution

Worldwide Distribution: USA and countries including: Australia, Belgium, Brazil, China, Czech Republic, Finland, France, Germany, India, Ireland, Italy, Japan, Kenya, Malaysia, Netherlands, Norway, Portugal, South Korea, Spain, Sweden, Switzerland, Taiwan, United Arab Emirates, United Kingdom.

Quantity

196 units both Transducer and ACUSon SC2000