FDA Recall Terminated

Ingenia MR-OR; 781437, 781438 The MR-OR option is intended for allowing transfer of a neurosurgical patient from OR suite to MR suite and back with tools and instructions for proper positioning and MR imaging of those patients. MR-OR option also facilitates execution of simple neurosurgical procedures in the near proximity of the MR scanner without a need to move the patient back to the OR suite

Recall: Z-0825-2017 · Initiated December 1, 2016

Recall

Recall Number
Z-0825-2017
Event Number
75844
Firm
Philips Electronics North America Corporation
FEI Number
1218950
Product Code
LNH
Status
Terminated
Root Cause
Device Design
Initiated
December 1, 2016
Posted
December 19, 2016
Terminated
August 31, 2017
Address
3000 Minuteman Rd, Andover, MA, 01810-1032

Description

Ingenia MR-OR; 781437, 781438 The MR-OR option is intended for allowing transfer of a neurosurgical patient from OR suite to MR suite and back with tools and instructions for proper positioning and MR imaging of those patients. MR-OR option also facilitates execution of simple neurosurgical procedures in the near proximity of the MR scanner without a need to move the patient back to the OR suite

Reason

Possibility that a patient or user may be exposed to a very small level of touch current.

Action

Philips sent an Urgent Field Safety Notice Medical Device Correction letter dated November 2016 to all affected customers. The letter explains the issue and actions planned by Philips. There is no need for action to be taken by customer or user. Customers with questions should contact their local Philips representative Technical Support Line 1-800-722-9377.

Distribution

Worldwide Distribution -US including CA, NY and Internationally to United Kingdom, Australia, Norway, and Thailand

Quantity

9