Ingenia MR-OR; 781437, 781438 The MR-OR option is intended for allowing transfer of a neurosurgical patient from OR suite to MR suite and back with tools and instructions for proper positioning and MR imaging of those patients. MR-OR option also facilitates execution of simple neurosurgical procedures in the near proximity of the MR scanner without a need to move the patient back to the OR suite
Recall
- Recall Number
- Z-0825-2017
- Event Number
- 75844
- Firm
- Philips Electronics North America Corporation
- FEI Number
- 1218950
- Product Code
- LNH
- Status
- Terminated
- Root Cause
- Device Design
- Initiated
- December 1, 2016
- Posted
- December 19, 2016
- Terminated
- August 31, 2017
- Address
- 3000 Minuteman Rd, Andover, MA, 01810-1032
Description
Ingenia MR-OR; 781437, 781438 The MR-OR option is intended for allowing transfer of a neurosurgical patient from OR suite to MR suite and back with tools and instructions for proper positioning and MR imaging of those patients. MR-OR option also facilitates execution of simple neurosurgical procedures in the near proximity of the MR scanner without a need to move the patient back to the OR suite
Possibility that a patient or user may be exposed to a very small level of touch current.
Philips sent an Urgent Field Safety Notice Medical Device Correction letter dated November 2016 to all affected customers. The letter explains the issue and actions planned by Philips. There is no need for action to be taken by customer or user. Customers with questions should contact their local Philips representative Technical Support Line 1-800-722-9377.
Worldwide Distribution -US including CA, NY and Internationally to United Kingdom, Australia, Norway, and Thailand
9