FDA Recall Terminated

Hobbs Medical 3 Prong Looped Retriever, a disposable device to be used by physician to remove foreign object or cauterize polyp specimens from the GI tract. REF: 5030 and 5030S

Recall: Z-0824-06 · Initiated March 15, 2006

Recall

Recall Number
Z-0824-06
Event Number
35011
Firm
Hobbs Medical, Inc.
FEI Number
1220592
Product Code
KOG
Status
Terminated
Root Cause
Process control
Initiated
March 15, 2006
Posted
May 3, 2006
Terminated
June 27, 2006
Address
8 Spring St, Stafford Springs, CT, 06076-1505

Description

Hobbs Medical 3 Prong Looped Retriever, a disposable device to be used by physician to remove foreign object or cauterize polyp specimens from the GI tract. REF: 5030 and 5030S

Reason

Diameter of device at the distal end would prevent it from passing through the endoscope's working channel.

Action

Hobbs Medical notfied customers by Registered Letter on 3/15/06, requesting return of recalled product.

Distribution

AK, IA, KY, MI, NH, NC,UT, WI, Canada, and UK

Quantity

254 units