FDA Recall Terminated

Finish Thompson Centrifugal Pump Model #110192-3 (sold as spare part)

Recall: Z-0823-2020 · Initiated April 19, 2019

Recall

Recall Number
Z-0823-2020
Event Number
83990
Firm
Isopure Corp
FEI Number
3013942511
Product Code
N/A
Status
Terminated
Root Cause
Nonconforming Material/Component
Initiated
April 19, 2019
Posted
January 19, 2020
Terminated
May 29, 2020
Address
11851 Plantside Dr, Louisville, KY, 40299-6328

Description

Finish Thompson Centrifugal Pump Model #110192-3 (sold as spare part)

Reason

There is a potential for an uneven plastic encapsulation of the magnet ring during the molding process and partial axial displacement. This issue could cause uneven wear during use, cause exposure of the magnet to solution, and thereby result in corrosion and the presence of iron.

Action

The only affected customer was contacted via telephone on 4/19/19. An URGENT PRODUCT FIELD CORRECTION NOTICE dated 5/22/19 was distributed to affected clinics.

Distribution

The products were distributed to the following US states: AL, FL, GA, MA, MD, MN, NC, NM, NY, OH, OK, TX, and VA.

Quantity

8