FDA Recall Terminated

This reagent is for in vitro diagnostic use. Pax-5 (1EW) monoclonal antibody is intended to be used for the qualitative identification by light microscopy of human Pax-5 in formalin-fixed, paraffin-embedded tissue by immunohistochemical staining using the automated Bond system.

Recall: Z-0823-2016 · Initiated December 30, 2015

Recall

Recall Number
Z-0823-2016
Event Number
73095
Firm
Leica Microsystems, Inc.
FEI Number
3000210133
Product Code
NJT
Status
Terminated
Root Cause
Component change control
Initiated
December 30, 2015
Posted
February 19, 2016
Terminated
January 5, 2017
Address
1700 Leider Ln, Buffalo Grove, IL, 60089-6622

Description

This reagent is for in vitro diagnostic use. Pax-5 (1EW) monoclonal antibody is intended to be used for the qualitative identification by light microscopy of human Pax-5 in formalin-fixed, paraffin-embedded tissue by immunohistochemical staining using the automated Bond system.

Reason

The Bond" Polymer Refine Detection and Novolink" Polymer Detection System may not provide adequate staining when using a detection protocol where one of its components - Peroxide Block / Peroxidase Block is applied after the primary marker incubation as specified in the Instructions for Use Following the Instructions for Use.

Action

The firm sent out a Medical Device Recall letter dated December 29th 2015 to two-hundred (200) consignees in the United States decribing the recall and the neccesary actions to be taken. These included: - Do not use or continue to use the listed antibody with the specified detection kits. - Order replacement detection kits for use with the listed antibody. - There is no need for retrospective review of tissue staining if appropriate positive controls were properly used and interpreted during the immuno / in-situ staining. -pass this notice primarily to the end users where the product has been sold and to all those within your organization who need to be aware of this issue. -Confirm reciept of the Recall Notice letter by signing and dating the attached Acknowledgement Form and faxing it back to us at 1-847-236-3747, or scan and email it to [email protected] Customer are requested to contact their local Leica representative if you have any questions or concerns, or alternatively please contact us at the following email address: [email protected] Peter Lloyd Leica Biosystems Newcastle Ltd Balliol Business Park West Benton Lane Newcastle upon Tyne NE 12 8EW United Kingdom Telephone +44 191 215 0567 Facsimile: +44 191 215 1152

Distribution

US Consignees incluude: AR ,AZ ,CA ,CO ,CT ,DC ,DE ,FL ,GA ,HI ,IA ,IL ,KY ,LA ,MA ,MD ,MI ,MN ,MO ,MT ,NC ,NH ,NJ ,NY ,OH ,OK ,PA ,PR ,SC ,TN ,TX ,UT ,VA ,VT ,WA ,WI and WV. Foreign Consignees include:Argentina , Australia , Brazil , Bulgaria , Chile , China , Ecuador ,Egypt , El Salvador , Finland , France , Georgia , Germany , Greece , India , Italy , Japan , Jordan , Kuwait , Lebanon , Malaysia , Philippines , Poland , Portugal , Romania , Russian Fed. , Singapore , South Korea , Spain , Sweden , Switzerland , Thailand , Tunisia , Turkey , United Kingdom and United Arab Emirates.

Quantity

1700 units