Femoral Shaft L=120mm; Femoral Shaft L=135mm; Femoral Shaft L=150mm The product is a component is a systems that is intended for the replacement if diseased or deficient bone around the knee joint, distal femur, proximal femur or total femur bone
Recall
- Recall Number
- Z-0822-2017
- Event Number
- 75892
- Firm
- Stanmore Implants Worldwide Ltd.
- FEI Number
- 3004105610
- Product Code
- KRO
- Status
- Terminated
- Root Cause
- Packaging
- Initiated
- December 7, 2016
- Posted
- December 19, 2016
- Terminated
- July 17, 2017
- Address
- 210 Centennial Avenue, Centennial Park Centennial Park Borehamwood United Kingdom
Description
Femoral Shaft L=120mm; Femoral Shaft L=135mm; Femoral Shaft L=150mm The product is a component is a systems that is intended for the replacement if diseased or deficient bone around the knee joint, distal femur, proximal femur or total femur bone
The recalling firm has received 5 customer inquiries reporting that at least one pouch of the double pouch packaging configuration has been punctured by the tip of the device's stem.
Stanmore Implants sent an Urgent Field Safety Notice dated November 25, 2016, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were advised to immediately check their internal inventory and quarantine all subject devices, circulate the Field Safety Notice internally to all interested/affected parties, maintain awareness of the notice internally until all required actions have been completed within their facility, inform Stanmore Implants if any of the subject devices have been distributed to other organizations, complete the attached customer response form and list of affected batches form, return the completed form and any affected devices to their S/W Representative. Customers with questions were instructed to contact their local sales representative.
Nationwide Distribution to OH, NE, FL, LA, NY, TX, MS, MD, CT, AL, PA
11 units