ECAT PET Scanners (mobile systems), distributed by Siemens Medical Solutions ECAT PET Scanners (mobile systems), and Reveal XL (mobile systems), distributed by CTI Molecular Imagining, Inc. Sceptre and Sceptre XL (mobile systems) distributed by Hitachi Medical Systems America
Recall: Z-0819-05
·
Initiated April 5, 2005
Recall
- Recall Number
- Z-0819-05
- Event Number
- 31770
- Firm
- CTI PET Systems Inc
- FEI Number
- 1034973
- Product Code
- KPS
- Status
- Terminated
- Root Cause
- Other
- Initiated
- April 5, 2005
- Posted
- May 11, 2005
- Terminated
- September 11, 2009
- Address
- 810 Innovation Drive, Knoxville, TN, 37932-2571
Description
ECAT PET Scanners (mobile systems), distributed by Siemens Medical Solutions ECAT PET Scanners (mobile systems), and Reveal XL (mobile systems), distributed by CTI Molecular Imagining, Inc. Sceptre and Sceptre XL (mobile systems) distributed by Hitachi Medical Systems America
Reason
Failure to properly secure the Mobile Bed Supports during transit could cause the support screws on the Patient Handling System to become loose and possibly break due to increased vibration during transit.
Action
Written notification via certified US Mail was sent out beginning 04/05/2005 to distributors and end users (with whom CTI had a service contract) to advise them of the problem.
Distribution
Nationwide, Canada, the Netherlands, Ireland & Denmark
Quantity
80 units