Foam electrodes with Silver-Carbon construction: P/N 202-9271, Cat No. 400-898, 2" x 2" Electrode, Foam, AG, All Serial Numbers distributed between 11/01/2009 and 01/31/2010 P/N 202-9281, Cat No. 202-9281, 2" Round Electrode, Foam, AG, All Serial Numbers distributed between 11/01/2009 and 01/31/2010 P/N 202-9251, Cat No. 400-893, 3" Round Electrode, Foam AG, All Serial Numbers distributed between 11/01/2009 and 01/31/2010 P/N 202-9291, Cat No. 400-892, 3" Round Electrode, Foam AG, All Serial Numbers distributed between 11/01/2009 and 01/31/2010 P/N 202-9371, Cat No. 400-888, 2" x 3.5" Electrode, Foam AG, All Serial Numbers distributed between 11/01/2009 and 01/31/2010 P/N 202-9491, Cat No. 400-869, 1.25" Round Electrode, Foam AG, All Serial Numbers distributed between 11/01/2009 and 01/31/2010 P/N 202-9511, Cat No. 400-853, 3" x 5" Electrode, Foam AG, All Serial Numbers distributed between 11/01/2009 and 01/31/2010 P/N 300-1541, Cat No. 300-154, 2" x 2.25" Electrode, Foam, AG, All Serial Numbers distributed between 11/01/2009 and 01/31/2010 P/N 300-1561, Cat No. 300-156, 2" Round Electrode, Foam, AG, All Serial Numbers distributed between 11/01/2009 and 01/31/2010 P/N 300-1611, Cat No. 300-161, 4" x 2.25" Electrode, Foam, AG, All Serial Numbers distributed between 11/01/2009 and 01/31/2010, 22818 P/N 300-1631, Cat No. 300-163, 3" Round Electrode, Foam, AG, All Serial Numbers distributed between 11/01/2009 and 01/31/2010 P/N 400-8701, Cat No. 400-870, 2" x 4" Oval Electrode, Foam, AG, All Serial Numbers distributed between 11/01/2009 and 01/31/2010. Cutaneous electrode.

Recall: Z-0818-2013 · Initiated November 30, 2009

Recall

Recall Number
Z-0818-2013
Event Number
63927
Firm
NAImco Inc dba Richmar Inc
FEI Number
3004750362
Product Code
GXY
Status
Terminated
Root Cause
Unknown/Undetermined by firm
Initiated
November 30, 2009
Posted
February 14, 2013
Terminated
September 3, 2013
Address
4120 S Creek Rd, Chattanooga, TN, 37406-1021

Description

Foam electrodes with Silver-Carbon construction: P/N 202-9271, Cat No. 400-898, 2" x 2" Electrode, Foam, AG, All Serial Numbers distributed between 11/01/2009 and 01/31/2010 P/N 202-9281, Cat No. 202-9281, 2" Round Electrode, Foam, AG, All Serial Numbers distributed between 11/01/2009 and 01/31/2010 P/N 202-9251, Cat No. 400-893, 3" Round Electrode, Foam AG, All Serial Numbers distributed between 11/01/2009 and 01/31/2010 P/N 202-9291, Cat No. 400-892, 3" Round Electrode, Foam AG, All Serial Numbers distributed between 11/01/2009 and 01/31/2010 P/N 202-9371, Cat No. 400-888, 2" x 3.5" Electrode, Foam AG, All Serial Numbers distributed between 11/01/2009 and 01/31/2010 P/N 202-9491, Cat No. 400-869, 1.25" Round Electrode, Foam AG, All Serial Numbers distributed between 11/01/2009 and 01/31/2010 P/N 202-9511, Cat No. 400-853, 3" x 5" Electrode, Foam AG, All Serial Numbers distributed between 11/01/2009 and 01/31/2010 P/N 300-1541, Cat No. 300-154, 2" x 2.25" Electrode, Foam, AG, All Serial Numbers distributed between 11/01/2009 and 01/31/2010 P/N 300-1561, Cat No. 300-156, 2" Round Electrode, Foam, AG, All Serial Numbers distributed between 11/01/2009 and 01/31/2010 P/N 300-1611, Cat No. 300-161, 4" x 2.25" Electrode, Foam, AG, All Serial Numbers distributed between 11/01/2009 and 01/31/2010, 22818 P/N 300-1631, Cat No. 300-163, 3" Round Electrode, Foam, AG, All Serial Numbers distributed between 11/01/2009 and 01/31/2010 P/N 400-8701, Cat No. 400-870, 2" x 4" Oval Electrode, Foam, AG, All Serial Numbers distributed between 11/01/2009 and 01/31/2010. Cutaneous electrode.

Reason

Electrode Socket Terminal pulling away from the electrode lead wire when the electrical stim lead wire was removed.

Action

Rich-Mar sent "URGENT: MEDICAL DEVICE VOLUNTARY RECALL" letters to affected customers beginning on 11/30/2009. The letter identified the product, problem, and actions to be taken by the customers. Contact the firm at 423-648-7730 ext. 214 for questions relating to t his recall.

Distribution

Distributed in the states of GA and TN.

Quantity

56980 electrodes