FDA Recall Terminated

VariSource AL 13197000, Tongue/Soft Tissue Kit

Recall: Z-0818-06 · Initiated February 21, 2006

Recall

Recall Number
Z-0818-06
Event Number
34969
Firm
Alpha Omega Services Inc
FEI Number
2022694
Product Code
JAQ
Status
Terminated
Root Cause
Other
Initiated
February 21, 2006
Posted
April 27, 2006
Terminated
August 30, 2006
Address
9156 Rose St, Bellflower, CA, 90706

Description

VariSource AL 13197000, Tongue/Soft Tissue Kit

Reason

There is a possibility that the diameter of the catheter does not meet specifications. As a result it may not be possible to correctly connect the catheter to a VariSource Quick Connect. This would prevent any treatment. Furthermore, since the problem would not be apparent until after the catheter has been implanted an unnecessary surgical implantation procedure would have taken place.

Action

Recall Notification Letter: began February 22, 2006. Depth/Scope: 100% of the customer''s who purchased the affected products will be identified and contacted with a faxed and mailed Medical Device Recall notice and followed up with a phone call by the AOS Sales & Marketing personnel. All customers will be requested return all affected products, both used and unused.

Distribution

Nationwide and Argentina, Brasil, Canada, Israel, Korea, , Taiwan, United Kingdom & Venezuela

Quantity

64