ARCHTECT System RS-232 Manual, product list number 06F71-04, for the ARCHITECT i2000 Processing Module and the ARCHITECT c8000 Processing Module. The product is manufactured and distributed by Abbott Laboratories Diagnostic Division, Irving, TX 75038
Recall: Z-0818-04
·
Initiated March 19, 2004
Recall
- Recall Number
- Z-0818-04
- Event Number
- 28728
- Firm
- Abbott Laboratories, Inc
- FEI Number
- 1628664
- Product Code
- JJE
- Status
- Terminated
- Root Cause
- Other
- Initiated
- March 19, 2004
- Posted
- July 20, 2004
- Terminated
- November 3, 2005
- Address
- 1921 Hurd Drive PO Box 152020, Irving, TX, 75015-2020
Description
ARCHTECT System RS-232 Manual, product list number 06F71-04, for the ARCHITECT i2000 Processing Module and the ARCHITECT c8000 Processing Module. The product is manufactured and distributed by Abbott Laboratories Diagnostic Division, Irving, TX 75038
Reason
When using a LIS the software system will send up a test result of up to 20 characters although the field is set to display only 15 characters per the RS-232 Manual.
Action
Customers were notified by letter on March 23, 2004 to contact their LIS vendor and to reconfigure the ASTM Field 10.1.4 to allow a maximum character length of 20 characters.
Distribution
Nationwide and to the following foreign countries: Chile, Canada, Germany, Taiwan, Australia, Signapore
Quantity
101