FDA Recall Terminated

Foam Electrodes with Carbon construction: P/N 201-1261, Cat Nos.: 201-126, 203-678, and 400-856, 1.5" x 3.5" Electrode, Foam, All Serial Numbers distributed between 11/01/2009 and 01/31/2010 P/N 201-1271, Cat Nos.: 201-127, 202-093, 203-483, 203-565, 203-673, and 400-873, 2" x 2" Electrode, Foam, All Serial Numbers distributed between 11/01/2009 and 01/31/2010 P/N 201-1281, Cat Nos.: 201-128, 202-277, 203-675, and 400-873, 2" Round Electrode, Foam, All Serial Numbers distributed between 11/01/2009 and 01/31/2010 P/N 201-1291, Cat Nos.: 201-129, 203-677, and 400-882, 3" Round Electrode, Foam, All Serial Numbers distributed between 11/01/2009 and 01/31/2010 P/N 201-1301, Cat Nos.: 201-130, 203-679, and 400-855, 1.5" x 3.5" Electrode, Foam, All Serial Numbers distributed between 11/01/2009 and 01/31/2010 P/N 201-1371, Cat Nos.: 201-137, 202-278, 203-484, 203-681, and 400-880, 2" x 3.5" Electrode, Foam, All Serial Numbers distributed between 11/01/2009 and 01/31/2010 P/N 202-5671, Cat No. 202-567, 2" x 2" Electrode, Foam, All Serial Numbers distributed between 11/01/2009 and 01/31/2010. Cutaneous electrode.

Recall: Z-0817-2013 · Initiated November 30, 2009

Recall

Recall Number
Z-0817-2013
Event Number
63927
Firm
NAImco Inc dba Richmar Inc
FEI Number
3004750362
Product Code
GXY
Status
Terminated
Root Cause
Unknown/Undetermined by firm
Initiated
November 30, 2009
Posted
February 14, 2013
Terminated
September 3, 2013
Address
4120 S Creek Rd, Chattanooga, TN, 37406-1021

Description

Foam Electrodes with Carbon construction: P/N 201-1261, Cat Nos.: 201-126, 203-678, and 400-856, 1.5" x 3.5" Electrode, Foam, All Serial Numbers distributed between 11/01/2009 and 01/31/2010 P/N 201-1271, Cat Nos.: 201-127, 202-093, 203-483, 203-565, 203-673, and 400-873, 2" x 2" Electrode, Foam, All Serial Numbers distributed between 11/01/2009 and 01/31/2010 P/N 201-1281, Cat Nos.: 201-128, 202-277, 203-675, and 400-873, 2" Round Electrode, Foam, All Serial Numbers distributed between 11/01/2009 and 01/31/2010 P/N 201-1291, Cat Nos.: 201-129, 203-677, and 400-882, 3" Round Electrode, Foam, All Serial Numbers distributed between 11/01/2009 and 01/31/2010 P/N 201-1301, Cat Nos.: 201-130, 203-679, and 400-855, 1.5" x 3.5" Electrode, Foam, All Serial Numbers distributed between 11/01/2009 and 01/31/2010 P/N 201-1371, Cat Nos.: 201-137, 202-278, 203-484, 203-681, and 400-880, 2" x 3.5" Electrode, Foam, All Serial Numbers distributed between 11/01/2009 and 01/31/2010 P/N 202-5671, Cat No. 202-567, 2" x 2" Electrode, Foam, All Serial Numbers distributed between 11/01/2009 and 01/31/2010. Cutaneous electrode.

Reason

Electrode Socket Terminal pulling away from the electrode lead wire when the electrical stim lead wire was removed.

Action

Rich-Mar sent "URGENT: MEDICAL DEVICE VOLUNTARY RECALL" letters to affected customers beginning on 11/30/2009. The letter identified the product, problem, and actions to be taken by the customers. Contact the firm at 423-648-7730 ext. 214 for questions relating to t his recall.

Distribution

Distributed in the states of GA and TN.

Quantity

129400 electrodes