Cloth electrodes with Silver-Carbon construction: P/N 202-9231, Cat No. 400-899, 2" x 2" Electrode, Cloth, AG, All Serial Numbers distributed between 11/01/2009 and 01/31/2010 P/N 202-9241, Cat No. 400-895, 2" Round Electrode, Cloth, AG, All Serial Numbers distributed between 11/01/2009 and 01/31/2010 P/N 202-9381, Cat No. 400-889, 2" x 3.5" Electrode, Cloth AG, All Serial Numbers distributed between 11/01/2009 and 01/31/2010 P/N 202-9501, Cat No. 202-9501, 1.25" Round Electrode, Cloth AG, All Serial Numbers distributed between 11/01/2009 and 01/31/2010 P/N 400-8991, Cat Nos.: 400-877T and 400-899T, 2" x 2.25" Electrode, Cloth, AG, All Serial Numbers distributed between 11/01/2009 and 01/31/2010. Cutaneous electrode.
Recall
- Recall Number
- Z-0816-2013
- Event Number
- 63927
- Firm
- NAImco Inc dba Richmar Inc
- FEI Number
- 3004750362
- Product Code
- GXY
- Status
- Terminated
- Root Cause
- Unknown/Undetermined by firm
- Initiated
- November 30, 2009
- Posted
- February 14, 2013
- Terminated
- September 3, 2013
- Address
- 4120 S Creek Rd, Chattanooga, TN, 37406-1021
Description
Cloth electrodes with Silver-Carbon construction: P/N 202-9231, Cat No. 400-899, 2" x 2" Electrode, Cloth, AG, All Serial Numbers distributed between 11/01/2009 and 01/31/2010 P/N 202-9241, Cat No. 400-895, 2" Round Electrode, Cloth, AG, All Serial Numbers distributed between 11/01/2009 and 01/31/2010 P/N 202-9381, Cat No. 400-889, 2" x 3.5" Electrode, Cloth AG, All Serial Numbers distributed between 11/01/2009 and 01/31/2010 P/N 202-9501, Cat No. 202-9501, 1.25" Round Electrode, Cloth AG, All Serial Numbers distributed between 11/01/2009 and 01/31/2010 P/N 400-8991, Cat Nos.: 400-877T and 400-899T, 2" x 2.25" Electrode, Cloth, AG, All Serial Numbers distributed between 11/01/2009 and 01/31/2010. Cutaneous electrode.
Electrode Socket Terminal pulling away from the electrode lead wire when the electrical stim lead wire was removed.
Rich-Mar sent "URGENT: MEDICAL DEVICE VOLUNTARY RECALL" letters to affected customers beginning on 11/30/2009. The letter identified the product, problem, and actions to be taken by the customers. Contact the firm at 423-648-7730 ext. 214 for questions relating to t his recall.
Distributed in the states of GA and TN.
25860 electrodes