FDA Recall Terminated

Cranex Tome panoramic x-ray unit with Spiral Tomography for cross-sectional imaging. manufactured by SOREDEX, PO Box 250, 00031 Helsinki, Finland.

Recall: Z-0816-2007 · Initiated April 28, 2006

Recall

Recall Number
Z-0816-2007
Event Number
37838
Firm
Instrumentarium Dental, Inc.
FEI Number
3005383085
Product Code
EDH
Status
Terminated
Root Cause
Other
Initiated
April 28, 2006
Posted
May 18, 2007
Terminated
February 24, 2008
Address
300 West Edgerton Avenue, Milwaukee, WI, 53207

Description

Cranex Tome panoramic x-ray unit with Spiral Tomography for cross-sectional imaging. manufactured by SOREDEX, PO Box 250, 00031 Helsinki, Finland.

Reason

There may be a material defect in the up/down movement axle of the unit. This may cause the upper part of the unit to suddenly come down.

Action

A recall letter, dated 04/28/2006, was sent to affected customers . The letter describes the issue, and requests that no patient operation is permitted until the unit is checked by authorized service technician. If the service technician noted wear of the screw, the assembly was replaced.

Distribution

Worldwide, including USA, Canada, Bermuda, and Colombia.

Quantity

9