Tube Stand for the Dynarad Phantom Mobile X-Ray System; Del Medical Systems, 11550 West King Street, Franklin Park, IL 60131

Recall: Z-0816-05 · Initiated March 10, 2005

Recall

Recall Number
Z-0816-05
Event Number
31456
Firm
Del Medical Systems Group
FEI Number
1418964
Product Code
IYB
Status
Terminated
Root Cause
Other
Initiated
March 10, 2005
Posted
May 11, 2005
Terminated
November 14, 2005
Address
11550 West King Street, Franklin Park, IL, 60131-1330

Description

Tube Stand for the Dynarad Phantom Mobile X-Ray System; Del Medical Systems, 11550 West King Street, Franklin Park, IL 60131

Reason

The Phantom Portable X-Ray System Stand may have a suspect weld supporting the bracket holding the locking cylinder assembly. It the weld fails, the tubehead arm assembly could fall and contact the patient or operator.

Action

Del Medical sent Field Correction Recall letters dated 3/10/05 to the direct accounts, informing them of the potential for a defective weld supporting the braket holding the locking cylinder assembly. If the weld fails, the tubehead arm assembly could potentially fall and contact the patient or operator. The accounts were requested to locate the affected systems and take them out of service until they are inspected and, if needed, repaired.

Distribution

Nationwide, Mexico and Thailand.

Quantity

31 stands