FDA Recall Open, Classified

GE Healthcare Optima XR220amx, Mobile Digital Radiographic System, Model Numbers: 1) 5555000-12; 2) 5555000-5; 3) 5555000-6; 4) A0659PB; 5) to be provided

Recall: Z-0812-2025 · Initiated November 25, 2024

Recall

Recall Number
Z-0812-2025
Event Number
95958
Firm
GE Medical Systems, LLC
FEI Number
2126677
Product Code
IZL
Status
Open, Classified
Root Cause
Component design/selection
Initiated
November 25, 2024
Posted
January 7, 2025
Address
3000 N Grandview Blvd, Waukesha, WI, 53188-1615

Description

GE Healthcare Optima XR220amx, Mobile Digital Radiographic System, Model Numbers: 1) 5555000-12; 2) 5555000-5; 3) 5555000-6; 4) A0659PB; 5) to be provided

Reason

GE HealthCare has become aware that in certain Optima XR200/220/240amx systems that have had a column replacement performed, some bolts used in the Column Safety mechanism can be missing. If these bolts are missing and a failure of the internal counterpoise cable occurs, the Column Safety mechanism will not work properly causing the arm with X-ray tube and collimator to drop downward on the column track.

Action

GE Healthcare issued an URGENT MEDICAL DEVICE CORRECTION notice to its consignees on 11/25/2024 via letter issued using traceable means. The notice explained the issue, safety issue, and requested the following actions be taken by the user: "1. Perform a check of your device prior to each patient exam to look for the following behaviors. If any are observed, DO NOT use the device and call GE HealthCare Service Immediately. . Column vertical movement requires increased effort. . Column has change in resistance (i.e. movement is not smooth) during vertical movement. . Column makes abnormal noises during vertical movement. 2. Please ensure all potential users in your facility are made aware of this safety notification and the recommended actions. 3. Please retain this document for your records. 4. Please complete and return the attached acknowledgement form to [email protected]" For questions or concerns regarding this notification, please contact GE HealthCare Service at 1-800-437-1171 or your local Service Representative. On 02/11/2025, the firm send an updated notice with a corrected GTIN for Optima XR200amx.

Distribution

Worldwide

Quantity

281 units