FDA Recall Terminated

NxStage Dialysate Sack: 1) REF SAK-301 UDI-M535SAK3010, 2) REF SAK-302 UDI-M535SAK3020, 3) REF SAK-303 UDI-M535SAK3030, 4) REF SAK-304 UDI-M535SAK3040, 5) REF SAK-306 UDI-M535SAK3060, 6) REF SAK-307 UDI- M535SAK3070, 6) REF SAK-402 UDI-M535SAK4020, 7) REF SAK-405 UDI- M535SAK4050, 8) REF SAK-406 UDI-M535SAK4060 & REF SAK-407 UDI- M535SAK4070 Hemodialysis systems and accessories

Recall: Z-0811-2018 · Initiated January 29, 2018

Recall

Recall Number
Z-0811-2018
Event Number
79185
Firm
NxStage Medical, Inc.
FEI Number
3003464075
Product Code
FKR
Status
Terminated
Root Cause
Material/Component Contamination
Initiated
January 29, 2018
Terminated
April 3, 2020
Address
350 Merrimack Street, Lawrence, MA, 01843-1748

Description

NxStage Dialysate Sack: 1) REF SAK-301 UDI-M535SAK3010, 2) REF SAK-302 UDI-M535SAK3020, 3) REF SAK-303 UDI-M535SAK3030, 4) REF SAK-304 UDI-M535SAK3040, 5) REF SAK-306 UDI-M535SAK3060, 6) REF SAK-307 UDI- M535SAK3070, 6) REF SAK-402 UDI-M535SAK4020, 7) REF SAK-405 UDI- M535SAK4050, 8) REF SAK-406 UDI-M535SAK4060 & REF SAK-407 UDI- M535SAK4070 Hemodialysis systems and accessories

Reason

May contain endotoxin levels which have been confirmed to exceed the ANSI/AAMI quality standard dialysis fluid of 0.5 EU/ml when used to prepare product.

Action

On January 29, 2018 Product Recall Notices & Recall Reply Forms included were emailed to those patients for whom the firm had an email address. Hard copy letters were sent via UPS second day air to patients for whom NxStage did not have an email address. All Dialysis centers were contacted via email. In addition to the active patients and dialysis center consignees of affected product, recall letters were also emailed to the Dialysis Center Home Training Nurse Manager for all active or dropped patients who were shipped affected product, and corporate contacts for national dialysis accounts. Customers are advised to inspect inventory, quarantine, discontinue use and return product. Customers are also encouraged to return the completed Recall Reply Form. Contact NxStage Customer Service to arrange for return of all affected product and for replacement product to be sent. Customers with questions may call Customer Service at 1-866-697-8243.

Distribution

Worldwide Distribution - USA (nationwide) and to the countries of : United Kingdom & Spain

Quantity

16,039 cases/2 units