Xhibit Telemetry Receiver, Model 96280, with Service Manual. Intended to provide monitoring system with patient data.
Recall
- Recall Number
- Z-0809-2025
- Event Number
- 95927
- Firm
- Spacelabs Healthcare, Inc.
- FEI Number
- 3010157426
- Product Code
- MHX
- Status
- Open, Classified
- Root Cause
- Labeling design
- Initiated
- November 25, 2024
- Posted
- January 7, 2025
- Address
- 35301 Se Center St, Snoqualmie, WA, 98065-9216
Description
Xhibit Telemetry Receiver, Model 96280, with Service Manual. Intended to provide monitoring system with patient data.
Telemetry receiver systems may experience a Windows proxy Issue causing the system to shut down and restart (Offline displayed), lasting approximately 1 to 1.5 minutes, which will cause temporary loss of monitoring. If a clinical event occurred during this period, staff may not detect until the system restarts; necessary clinical intervention could be delayed.
On 11/25/2024, correction notices were sent to customers asking them to do the following: 1) Service manual has been updated by adding a section for "Required periodic system maintenance of Spacelabs Telemetry Farm", pages 5-30 through 5-35. 2) Download current versions of the Service Manual and PM Instructions at https://manuals.spacelabshealthcare.com/ 3) You may also obtain a copy of the latest manual by contacting the firm's Technical Support. 4) Discard previous copies of these documents 5) Complete the online customer acknowledgement form which can be accessed by using URL: https://spacelabshealthcare.com/support/recall-reply-form/recall-reply-form-xtr-windows-web-proxy-issue/ For additional information or to report any adverse events as a result of this issue, please contact the firm's Technical Support 1-800-522-7025 and select 2 for Technical Support
US: OK, MN, IA, SC, NE, MI, NY, KY, GA, MO, MT, LA, WA, SD, CO, PA, WY, TX, ID, NJ, FL, OH, MS, OR, AK, IL, AR, KS, AL, TN, WI, VA, NM, CA, IN, RI, NV, MA, NC, ME, CT, AZ, MD, PR. OUS: Kuwait, Australia, Canada, Roumanie, Greece, Philippines, Spain, Italy, Hong Kong, Switzerland, Panama, Taiwan, Netherlands, Malaysia, Poland, Mexico, United Kingdom, Saudi Arabia, France, Czech Republic