Programming software (all versions) for Polymedco PolyTiter Immunofluorescent Titration System and used with DiaSorin ANAFLUOR Indirect Fluorescent Antibody Test.

Recall: Z-0807-05 · Initiated September 24, 2004

Recall

Recall Number
Z-0807-05
Event Number
30946
Firm
Diasorin Inc.
FEI Number
2182595
Product Code
DHN
Status
Terminated
Root Cause
Other
Initiated
September 24, 2004
Posted
May 10, 2005
Terminated
July 22, 2006
Address
1951 Northwestern Ave S, Stillwater, MN, 55082-7536

Description

Programming software (all versions) for Polymedco PolyTiter Immunofluorescent Titration System and used with DiaSorin ANAFLUOR Indirect Fluorescent Antibody Test.

Reason

A problem with software for the PolyTiter Immunofluorescent Titration System in that no flags or warnings appear when PolyTiter Calibrator Values are re-entered or changed (including values changed to unacceptable values) after the initial curve has been generated, if the 'refresh button' has not been pressed. The graph may appear valid; however erroneous results may occur.

Action

Customers were sent a Technical Bulletin describing the correct procedure to be used when entering data into the Poly/Titer tiration system, until the updated software is installed