FDA Recall Terminated

AdVance"Male Sling System, REF 720088-01 & AdVance" XP Male Sling System REF 720163-01 (not commercially available in the US) The AdVance Male Sling System consists of two, sterile single-use surgical instruments called helical needle passers and a mesh sling with attached connectors. One end of each helical needle passer is keyed to allow a secure attachment of the sling connectors. The sling mesh is constructed of polypropylene monofilament mesh that is precut to 1.2 centimeters arm width, 3.55 centimeters center width, and 35.5 centimeters length. Two absorbable tensioning sutures are threaded into the length of the mesh. Plastic sheaths are placed over each arm of the mesh to aide in ease of placement. The connectors are attached to the ends of the helical needle passers during the procedure. The mesh and suture portions of the sling are intended to remain in the body as a permanent implant and the mesh component is not absorbed or degraded by the action of tissue in-growth or tissue enzymes.

Recall: Z-0806-2015 · Initiated October 15, 2014

Recall

Recall Number
Z-0806-2015
Event Number
69796
Firm
American Medical Systems, Inc.
FEI Number
1000116179
Product Code
OTM
Status
Terminated
Root Cause
Nonconforming Material/Component
Initiated
October 15, 2014
Posted
December 16, 2014
Terminated
May 28, 2015
Address
10700 Bren Rd, W Minnetonka, MN, 55343-9679

Description

AdVance"Male Sling System, REF 720088-01 & AdVance" XP Male Sling System REF 720163-01 (not commercially available in the US) The AdVance Male Sling System consists of two, sterile single-use surgical instruments called helical needle passers and a mesh sling with attached connectors. One end of each helical needle passer is keyed to allow a secure attachment of the sling connectors. The sling mesh is constructed of polypropylene monofilament mesh that is precut to 1.2 centimeters arm width, 3.55 centimeters center width, and 35.5 centimeters length. Two absorbable tensioning sutures are threaded into the length of the mesh. Plastic sheaths are placed over each arm of the mesh to aide in ease of placement. The connectors are attached to the ends of the helical needle passers during the procedure. The mesh and suture portions of the sling are intended to remain in the body as a permanent implant and the mesh component is not absorbed or degraded by the action of tissue in-growth or tissue enzymes.

Reason

During routine periodic packaging testing, AMS identified that in simulated extreme shipping and handling situations the sterile barrier packaging of the Needle Passer components supplied with the AdVance" Male Sling System; Model # 720088-01; AdVance" XP Male Sling System; Model # 720163-01; Monarc" C Subfacial Hammock; Model # 72404195; Monarc" + Subfacial Hammock; Model #

Action

Consignees were sent an AMS "Urgent Recall Notice" dated October 15, 2014. The letter was addressed to Physician, Health Care Professional, Nurse and Risk Manger. The letter described the problem and the product involved in the recall. Requested consignees to return all affected product and to complete and return the Acknowledgement Form. For questions, they can contact American Medical Systems Customer Service at 1-800-328-3881 (x6469). A second letter was available as optional communication to implanting physicians known by AMS to be regular users of the device. This physician letter was dated October 20, 2014.

Distribution

Worldwide Distribution-US (nationwide) including DC and Puerto Rico and the countries of AUSTRIA, BELGIUM, CROATIA, CECH REPUBLIC, ESTONIA, FINLAND, FRANCE, GERMANY, GREECE, IRELAND, ISRAEL, ITALY, LEBANON, NETHERLANDS, NORWAY, POLAND, PORTUGAL, SAUDI ARABIA, SLOVENIA, SERBIA, SOUTH AFRICA, SPAIN. SWEDEN, SWITZERLAND, TURKEY, UNITED KINGDOM, CHILE, COLUMBIA, COSTA RICA, HONG KONG, MALAYSIA, MARTINIQUE, NEW ZEALAND, PANAMA, PHILIPPINES, SINGAPORE, THAILAND, VENEZUELA, CANADA, AUSTRALIA.

Quantity

10,085