FDA Recall Terminated

ORCHESTRA/ORCHESTRA PLUS Programmer

Recall: Z-0805-2014 · Initiated November 20, 2013

Recall

Recall Number
Z-0805-2014
Event Number
67013
Firm
Sorin Group Italia S.r.l. Via Crescentino Saluggia, VC Italy
FEI Number
1000165971
Product Code
LNX
Status
Terminated
Root Cause
Software design
Initiated
November 20, 2013
Posted
January 22, 2014
Terminated
September 8, 2014

Description

ORCHESTRA/ORCHESTRA PLUS Programmer

Reason

Sorin has voluntarily issued a notification to physicians related to the overestimation of the residual longevity displayed by its programmer (ORCHESTRA or ORCHESTRA PLUS) during a follow-up exam of patients implanted with a REPLY or ESPRIT pacemaker.

Action

On Nov 20 and Nov 25, 2013, the Dear Doctor Letter was sent by certified mail to 1,310 physicians who are following 13,003 patients implanted with a REPLU or ESPRIT pacemakers in the US. Sorin provided the following recommendations to the physicians: You should consider checking the battery impedance of the last follow-up exam. In case the battery impedance is greater than or equal to 3.5k, a follow-up visit must be scheduled within a maximum of 6 months from the last follow-up visit. When pacemaker operation is checked by the simple application of a magnet, a magnet rate less than 95 min-1 should trigger a follow-up exam in the pacemaker centre. As a general rule, a maximum of 6 month follow-up interval when the battery impedance becomes greater than or equal to 3.5 k. This recommendation should also be followed subsequent to the installation of the new programmer software version.

Distribution

Nationwide.

Quantity

1,718