GE LightSpeed Xtra / GE LightSpeed RT Pro 16, Model 2374681-6, CT Scanner systems.
Recall
- Recall Number
- Z-0805-2007
- Event Number
- 37007
- Firm
- GE Healthcare
- FEI Number
- 2126677
- Product Code
- JAK
- Status
- Terminated
- Root Cause
- Other
- Initiated
- December 21, 2006
- Posted
- May 15, 2007
- Terminated
- July 3, 2012
- Address
- 3000 N Grandview Blvd, Waukesha, WI, 53188-1615
Description
GE LightSpeed Xtra / GE LightSpeed RT Pro 16, Model 2374681-6, CT Scanner systems.
Oil may leak onto the port window of the CT tube and may result in image artifacts that could lead to misdiagnosis.
A Field Modification Instruction was issued to the field to inspect all systems and replace the tube window screws and the window filter, as well as clean any traces of oil in the tube window area. A 12/21/2006, Urgent Product Correction Notice/Customer letter has also been distributed informing the customers of the safety issue and the product correction.
Worldwide Distribution: USA in the following states and locations: AL, AK, AZ, AR, CA, CO, CT, DE, DC, FL, GA, HI, ID, IL, IN, IA, KS, KY, LA, ME, MD, MA, MI, MN, MS, MO, MT, NE, NV, NH, NJ, NY, NC, ND, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WV, WI, Guam and Puerto Rico. Foreign countries include ; Albania, Algeria, Argentina, Austria, Australia, Belgium, Bulgaria, Brazil, Belarus, Canada, Chile, China, Columbia, Czechoslovakia, Denmark, Finland, France, Germany, Great Britain, Greece, Honduras, Hong Kong, Iceland, India, Indonesia, Ireland, Israel, Italy, Japan, Korea, Kuwait, Latvia, Lebanon, Lithuania, Luxembourg, Malaysia, Martinique, Mexico, Netherlands, New Zealand, Norway, Panama, Philippines, Poland, Portugal, Reunion Island, Romania, Russia, Saudi Arabia, Singapore, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Tunisia, Turkey, United Arab Emirates, Venezuela, and Vietnam
21