FDA Recall Terminated

GE LightSpeed Pro 16 (100kW/80kW) Models 2357739, 2357739-3, 2357739-4, 2357739-6, 2357739-7 and 2357739-8, CT Scanner systems

Recall: Z-0804-2007 · Initiated December 21, 2006

Recall

Recall Number
Z-0804-2007
Event Number
37007
Firm
GE Healthcare
FEI Number
2126677
Product Code
JAK
Status
Terminated
Root Cause
Other
Initiated
December 21, 2006
Posted
May 15, 2007
Terminated
July 3, 2012
Address
3000 N Grandview Blvd, Waukesha, WI, 53188-1615

Description

GE LightSpeed Pro 16 (100kW/80kW) Models 2357739, 2357739-3, 2357739-4, 2357739-6, 2357739-7 and 2357739-8, CT Scanner systems

Reason

Oil may leak onto the port window of the CT tube and may result in image artifacts that could lead to misdiagnosis.

Action

A Field Modification Instruction was issued to the field to inspect all systems and replace the tube window screws and the window filter, as well as clean any traces of oil in the tube window area. A 12/21/2006, Urgent Product Correction Notice/Customer letter has also been distributed informing the customers of the safety issue and the product correction.

Distribution

Worldwide Distribution: USA in the following states and locations: AL, AK, AZ, AR, CA, CO, CT, DE, DC, FL, GA, HI, ID, IL, IN, IA, KS, KY, LA, ME, MD, MA, MI, MN, MS, MO, MT, NE, NV, NH, NJ, NY, NC, ND, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WV, WI, Guam and Puerto Rico. Foreign countries include ; Albania, Algeria, Argentina, Austria, Australia, Belgium, Bulgaria, Brazil, Belarus, Canada, Chile, China, Columbia, Czechoslovakia, Denmark, Finland, France, Germany, Great Britain, Greece, Honduras, Hong Kong, Iceland, India, Indonesia, Ireland, Israel, Italy, Japan, Korea, Kuwait, Latvia, Lebanon, Lithuania, Luxembourg, Malaysia, Martinique, Mexico, Netherlands, New Zealand, Norway, Panama, Philippines, Poland, Portugal, Reunion Island, Romania, Russia, Saudi Arabia, Singapore, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Tunisia, Turkey, United Arab Emirates, Venezuela, and Vietnam

Quantity

462