Civco Belly Board MRI, Reusable non-sterile glass fiber belly board with cushion, REF 126000, packaged 1 per box. Product Usage: This device is intended to reduce the irradiated small bowel volume of gynecological patients undergoing treatment in the pelvic region.
Recall
- Recall Number
- Z-0803-2014
- Event Number
- 66591
- Firm
- Civco Medical Instruments Inc
- FEI Number
- 1937223
- Product Code
- IYE
- Status
- Terminated
- Root Cause
- Device Design
- Initiated
- October 25, 2013
- Posted
- January 21, 2014
- Terminated
- February 2, 2017
- Address
- 102 1st St S, Kalona, IA, 52247-9589
Description
Civco Belly Board MRI, Reusable non-sterile glass fiber belly board with cushion, REF 126000, packaged 1 per box. Product Usage: This device is intended to reduce the irradiated small bowel volume of gynecological patients undergoing treatment in the pelvic region.
Scales adhered to the product may be misaligned from the zero reference such that one side will not match the other.
Civco Medical Instruments Inc sent a FIELD SAFETY NOTIFICATION letter dated October 28, 2013 via UPS to their customers. The notification identified the affected device, problem, and field safety corrective action. The letter describes how customers can check to see whether the scales on their device are out of alignment. The letter states CIVCO will rework the system to correct the difference. For questions contact sales representative at CIVCO Medical Solutions, Radiation Oncology at 800-842-8688 or 712-737-8688.
Worldwide Distribution - USA Nationwide in the states of AL, AR, AZ, CA, CO, CT, FL, GA, IL, KS, MA, MD, MI, MN, MT, NC, NJ, NY, OH, OK, PA, SC, TN, TX, VA, WA, WI, and WV. Foreign distribution was made to Brazil, Canada, Chile, Czech Republic, France, Germany, Iraq, Israel, Italy, Japan, Malaysia, Saudi Arabia, Slovenia, South Korea, Spain, Turkey, and United Arab Emirates.
2 belly board