FDA Recall
Terminated
Terumo CDI 101 Hematocrit/Oxygen Saturation Monitoring System. Monitors blood/oxygen saturation, hematocrit and hemoglobin levels.
Recall: Z-0802-2013
·
Initiated November 1, 2010
Recall
- Recall Number
- Z-0802-2013
- Event Number
- 60695
- Firm
- Terumo Cardiovascular Systems Corporation
- FEI Number
- 1828100
- Product Code
- DRY
- Status
- Terminated
- Root Cause
- Software design
- Initiated
- November 1, 2010
- Posted
- February 12, 2013
- Terminated
- November 8, 2013
- Address
- 6200 Jackson Road, Ann Arbor, MI, 48103-9586
Description
Terumo CDI 101 Hematocrit/Oxygen Saturation Monitoring System. Monitors blood/oxygen saturation, hematocrit and hemoglobin levels.
Reason
Inaccurate readings. Software contains unused (turned off) accuracy correction factors for slope (gain) and offset (bias) for Hematocrit/Saturation calculations. The accuracy on any cuvette size did not meet the specifications.
Action
Terumo sent an "URGENT MEDICAL DEVICE CORRECTION" letter dated November 1, 2010 to all affected customers. The letter identified the product, problem, and the actions to be taken by the customers. Contact Terumo CVS Customer Service at 1-800-521-2818 for questions or concerns.
Distribution
Worldwide Distribution.
Quantity
622 units