FDA Recall Terminated

Terumo CDI 101 Hematocrit/Oxygen Saturation Monitoring System. Monitors blood/oxygen saturation, hematocrit and hemoglobin levels.

Recall: Z-0802-2013 · Initiated November 1, 2010

Recall

Recall Number
Z-0802-2013
Event Number
60695
Firm
Terumo Cardiovascular Systems Corporation
FEI Number
1828100
Product Code
DRY
Status
Terminated
Root Cause
Software design
Initiated
November 1, 2010
Posted
February 12, 2013
Terminated
November 8, 2013
Address
6200 Jackson Road, Ann Arbor, MI, 48103-9586

Description

Terumo CDI 101 Hematocrit/Oxygen Saturation Monitoring System. Monitors blood/oxygen saturation, hematocrit and hemoglobin levels.

Reason

Inaccurate readings. Software contains unused (turned off) accuracy correction factors for slope (gain) and offset (bias) for Hematocrit/Saturation calculations. The accuracy on any cuvette size did not meet the specifications.

Action

Terumo sent an "URGENT MEDICAL DEVICE CORRECTION" letter dated November 1, 2010 to all affected customers. The letter identified the product, problem, and the actions to be taken by the customers. Contact Terumo CVS Customer Service at 1-800-521-2818 for questions or concerns.

Distribution

Worldwide Distribution.

Quantity

622 units