Computed Tomography X-Ray System (Brilliance CT Big Bore Oncology & Brilliance CT Big Bore Radiology) Distributed by Philips Healthcare System, Cleveland, OH. Produces cross-sectional body images by computer reconstruction of x-ray transmission data taken at different angles and planes.

Recall: Z-0801-2014 · Initiated December 4, 2013

Recall

Recall Number
Z-0801-2014
Event Number
67168
Firm
Philips Medical Systems (Cleveland) Inc
FEI Number
1525965
Product Code
JAK
Status
Terminated
Root Cause
Software design
Initiated
December 4, 2013
Posted
January 21, 2014
Terminated
July 1, 2015
Address
595 Miner Rd, Cleveland, OH, 44143-2131

Description

Computed Tomography X-Ray System (Brilliance CT Big Bore Oncology & Brilliance CT Big Bore Radiology) Distributed by Philips Healthcare System, Cleveland, OH. Produces cross-sectional body images by computer reconstruction of x-ray transmission data taken at different angles and planes.

Reason

Machine Name in TumorLOC Basic Mode is obscured. When TumorLOC is in Basic Mode and Set Vertical Laser equals Yes, the Machine Name label and field are covered by the controls above it. The Machine Name cannot be viewed or changed.

Action

On 12/4/2013 the firm sent Urgent Medical Device Corrections letters to their customers.

Distribution

Worldwide Distribution-USA (nationwide) including the states of AK, MD, NH, NV, OH, and OR, and the countries of Australia, China, France, India, Indonesia, Italy, Korea, Mexico, Poland, Romania, Russia, Spain, Switzerland, Taiwan, Thailand & United Arab Emirates.

Quantity

32 units