FDA Recall Terminated

Maxi Move Patient Lift with Scale; Manufactured by Arjo Hospital Equipment AB, Eslov, Sweden; Model KMBB4OSU2FUS

Recall: Z-0799-2007 · Initiated January 5, 2007

Recall

Recall Number
Z-0799-2007
Event Number
37064
Firm
Arjo, Inc.
FEI Number
1419652
Product Code
FSG
Status
Terminated
Root Cause
Other
Initiated
January 5, 2007
Posted
May 11, 2007
Terminated
September 2, 2008
Address
50 Gary Ave, Ste A, Roselle, IL, 60172-1684

Description

Maxi Move Patient Lift with Scale; Manufactured by Arjo Hospital Equipment AB, Eslov, Sweden; Model KMBB4OSU2FUS

Reason

The hanger bar assembly may detach from the Maxi Move lifters with scale, due to an incorrect torque setting on the two bolts holding the scale in place.

Action

Arjo sent Urgent Device Field Correction Recall Notice letters dated 1/5/07 to the end user accounts who received the affected lifts, advising them of the potential for the hanger bar assembly to detach. They were informed that an Arjo Field Service Technician would call them within five business days to schedule the time for him to visit and perform the inspection and correction to the lifts involved. If they did not hear from the technician within that time, they were instructed to contact the Arjo Quality Department at 800-323-1245, ext. 6140. The accounts were also requested to complete and return the enclosed customer response form, acknowledging the receipt and understanding of the letter.

Distribution

Nationwide

Quantity

121 lifts