FDA Recall Terminated

Deltec Cozmo Insulin Pump, Model 1700, Product Catalog Numbers 21-1711-01, 21-1712-01 and 21-1713-01. The 3 versions vary only in the color of the exterior.

Recall: Z-0798-03 · Initiated March 14, 2003

Recall

Recall Number
Z-0798-03
Event Number
25780
Firm
Deltec, Inc
FEI Number
2183502
Product Code
LZG
Status
Terminated
Root Cause
Other
Initiated
March 14, 2003
Posted
May 9, 2003
Terminated
July 2, 2004
Address
1265 Grey Fox Road, St Paul, MN, 55112

Description

Deltec Cozmo Insulin Pump, Model 1700, Product Catalog Numbers 21-1711-01, 21-1712-01 and 21-1713-01. The 3 versions vary only in the color of the exterior.

Reason

A software error could have resulted in the delivery of a larger bolus of insulin than anticipated when the pump was programmed in a particular manner.

Action

Recall Notification letters were sent to the consignees on March 14, 2003. The letters gave recommendations for avoiding the problem, stated that updated software to prevent the problem would be introduced, and requested that the customer call a toll-free phone number to arrange replacement of their affected pump.

Distribution

The product was distributed nationwide in the United States. There are no foreign or U.S. Government accounts.

Quantity

913 pumps